Application Note Number 2

An Optimum Test Sequence for Dynalyzers

Introduction:

The Dynalyzer x-ray calibration system is a highly accurate tool for the measurement of x-ray generator voltages and currents. To use this instrument efficiently, it is necessary to understand some of the physics behind the measurements. This application note will outline the procedure that will produce the maximum accuracy, and then will describe the physical reasoning behind it.

Optimal Test Protocol and Rationale:

The procedure outlined below will enable a technician to calibrate properly an x-ray generator, when combined with the protocol of the equipment manufacturer. This procedure must be modified for specific types of generators.

1) Equipment Installation:

Install the Dynalyzer between the x-ray tube and the x-ray generator according to Figure 5. Connect the Dynalyzer terminal exactly as shown. Connect both the anode and cathode cables.

2) Initial Adjustments:

Following the machine specific instructions, adjust the anode current for each tube station during tube installation, or during preventive maintenance, first measure the KV and mA for each station.

Sum of A+C

4% Ripple @

720 Hz
Notes 2 - Figure 6b.gif

Figure 6B

In a 6 pulse generator, the secondary coils for the positive and negative voltages, with respect to ground, are symmetrical, and are usually both Wye connected. This symmetry results in the anode and cathode waveforms to ground being in phase. In a 12 pulse generator, there are essentially two transformers. One has a delta primary and one has a wye primary, the two secondaries may be wye and have a full wave rectifier output producing 87.5 KVP to ground. There is a 30 degree phase shift between the anode and cathode ripple waveform resulting in the need for vector addition of the voltages, not algebraic addition of A (only) and C (only). Therefore it is necessary to have the voltage divider in both the anode and cathode during the initial set up of a three phase 12 pulse generator.

3) Final Adjustments

Once the generator is set up, the KVP will be a function of the KV compensation circuitry and the expected mA.

The divider or Dynalyzer may be removed from the cathode circuit, and the x-ray transformer cable from the cathode is directly attached from tube to transformer. Then the mA may be verified. If the mA has changed and the generator has a " master" filament adjustment or a "cable compensation adjustment," then adjust it get the correct mA value on that filament, and verify that the other mA stations have been brought back into adjustment. During this procedure, it is not necessary to verify or reset the KVP, however if the anode (only) data was previously recorded, it could be monitored.

Conclusion:

From the above discussion, the important points to remember when calibrating a x-ray generator are:

Note2Fig6.gif

Figure 6

Anode Voltage

Cathode Voltage

Notes 2 - Figure 6a.gif
Three Phase 12 Pulse Anode and Cathode Waveform
Notice that the peaks are not in phase.

Figure 6a